Measurement Summary Report
Transpose
Measurement Summary from 01/01/2020 00:00:00 HKT to 05/01/2020 00:00:00 HKT
Measurements
186017000
186015082
Test Compositional Accuracy - BD5%AU
(187
/
15
) 92.574%
Test Analog Output2 set to 0.0 VDC
(356
/
0
) 100%
Test Pump A Dynamic Leak Low PSI - Accum. Max Pressure
(356
/
0
) 100%
Test Flow Rate Accuracy Pump A&B 1.0ml+1.0ml/min
(356
/
0
) 100%
Test Seal Wash Pump Prime/Flow
(356
/
1
) 99.72%
Test Pump A Dynamic Leak - Primary Percent of Stroke
(356
/
0
) 100%
Test Pump B Dynamic Leak Low PSI - Primary Max Pressure
(356
/
3
) 99.164%
Test Compositional Accuracy - BD95%AU
(187
/
15
) 92.574%
Test Pump A Dynamic Leak Low PSI - Primary Percent of Stroke
(356
/
0
) 100%
Test Flow Rate Accuracy Pump A 0.5ml/min
(356
/
11
) 97.003%
Test Input C Prime with MeOH
(187
/
5
) 97.396%
Test Pump A Dynamic Leak - Accum. Max Pressure
(356
/
2
) 99.441%
Test Static Leak Decay VV to Mixer1 to VV
(187
/
2
) 98.942%
Test Pump B Dynamic Leak - Accum. Percent of Stroke
(356
/
0
) 100%
Test Pump B Dynamic Leak - Primary Percent of Stroke
(356
/
0
) 100%
Test Pump B Dynamic Leak Low PSI - Accum. Max Pressure
(356
/
0
) 100%
Test Pressure Ripple at 5.0 ml/min
(187
/
0
) 100%
Test Degasser Pressure
(356
/
0
) 100%
Test Analog Output1 set to 0.0 VDC
(356
/
0
) 100%
Test Pump A Input 2 Prime with IPA
(356
/
10
) 97.268%
Test Flow Rate Accuracy Pump B 0.5ml/min
(356
/
3
) 99.164%
Test Pump B Dynamic Leak Low PSI - Primary Percent of Stroke
(356
/
0
) 100%
Test Pump A Dynamic Leak - Primary Max Pressure
(356
/
9
) 97.534%
Test Pump A Pressure Ripple
(356
/
2
) 99.441%
Test Analog Output1 set to 2.0 VDC
(356
/
0
) 100%
Test Pump A Dynamic Leak Low PSI - Primary Leak Rate
(356
/
22
) 94.18%
Test Pump B Dynamic Leak Low PSI - Accum. Leak Rate
(356
/
3
) 99.164%
Test Input D Prime with MeOH
(187
/
13
) 93.5%
Test Compositional Accuracy - AC95%AU
(187
/
14
) 93.035%
Test High Pressure
(187
/
2
) 98.942%
Test Input A Prime with MeOH
(187
/
1
) 99.468%
Test Pump A Dynamic Leak - Accum. Percent of Stroke
(356
/
0
) 100%
Test Pump B Dynamic Leak - Primary Max Pressure
(356
/
5
) 98.615%
Test Input B Prime with MeOH
(187
/
6
) 96.891%
Test Pump B Dynamic Leak - Accum. Max Pressure
(356
/
4
) 98.889%
Test Static Leak Decay VV to Mixer2 to VV
(187
/
2
) 98.942%
Test Pump B Dynamic Leak Low PSI - Accum. Percent of Stroke
(356
/
0
) 100%
Test Dynamic Leak - Primary Compressed Volume
(187
/
7
) 96.392%
Test Dynamic Leak - Primary Maximum Pressure
(187
/
7
) 96.392%
Test Analog Output2 set to 2.0 VDC
(356
/
0
) 100%
Test Pump A Dynamic Leak - Accum. Leak Rate
(356
/
8
) 97.802%
Test Pump B Dynamic Leak Low PSI - Primary Leak Rate
(356
/
11
) 97.003%
Test Pump B Input 2 Prime with IPA
(356
/
5
) 98.615%
Test Pump A Dynamic Leak Low PSI - Accum. Leak Rate
(356
/
7
) 98.072%
Test Pump B Pressure Ripple
(356
/
3
) 99.164%
Test Dynamic Leak - Accumulator Leak Rate
(187
/
7
) 96.392%
Test Pump A Input 1 Prime with MeOH
(356
/
0
) 100%
Test Pump B Dynamic Leak - Accum. Leak Rate
(356
/
5
) 98.615%
Test Degasser Pressure
(187
/
1
) 99.468%
Test Dynamic Leak - Primary Leak Rate
(187
/
9
) 95.408%
Test Pump B Dynamic Leak - Primary Leak Rate
(356
/
6
) 98.343%
Test Pump A Dynamic Leak - Primary Leak Rate
(356
/
8
) 97.802%
Test High Pressure Flow Pump A&B
(356
/
0
) 100%
Test Dynamic Leak - Primary Percent of Stroke
(187
/
7
) 96.392%
Test Dynamic Leak - Accumulator Compressed Volume
(187
/
0
) 100%
Test Pump B Input 1 Prime with MeOH
(356
/
2
) 99.441%
Test Flow Rate Accuracy Pump B 2ml/min
(356
/
1
) 99.72%
Test Pump A Dynamic Leak Low PSI - Accum. Percent of Stroke
(356
/
0
) 100%
Test Pump A Dynamic Leak Low PSI - Primary Max Pressure
(356
/
4
) 98.889%
Test Seal Wash Pump Prime/Flow
(187
/
1
) 99.468%
Test Compositional Accuracy - AC5%AU
(187
/
15
) 92.574%
Test Flow Rate Accuracy Pump A 2ml/min
(356
/
7
) 98.072%
Test Flow Rate Accuracy 0.3ml/min
(187
/
19
) 90.777%
Test Dynamic Leak - Accumulator Percent of Stroke
(187
/
0
) 100%
Test Pressure Ripple at 2.5 ml/min
(187
/
0
) 100%
Test Dynamic Leak Manifold to Filter Tube
(356
/
30
) 92.228%
Test Dynamic Leak - Accum. Maximum Pressure
(187
/
0
) 100%
Test Flow Rate Accuracy 2.5ml/min
(187
/
3
) 98.421%