**************************************** UUT Report Serial Number: 200819009C Date: 11/7/2019 Time: 9:57:35 AM Operator: tester UUT Result: Failed **************************************** Fixture ID: Passed Fixture ID OK Fixture Down Detect: Passed Fixture Down Test Power Supplies: Passed Test Power: Passed Measurement: 5.00768 Limits: Low: 4.75 High: 5.25 Comparison Type: GELE (>= <=) Test Power: Passed Measurement: 12.0053 Limits: Low: 11.7 High: 12.3 Comparison Type: GELE (>= <=) Test Power: Passed Measurement: 12.0027 Limits: Low: 11.7 High: 12.3 Comparison Type: GELE (>= <=) Test Serial Communications: Failed Test Instrument Identification String: Failed String: '' Limits: String: Pump Control Module Comparison Type: Ignore Case Power On Initialization Sequence Failed End UUT Report